111
Y
(mg/dl)
X
(mg/dl)
FDC
reference
Y=1.1X-5
Y=X
0 10 600
0
600
A B
Measurement using your instruments
If the regression formula is Y = 1.1X - 5:
A = 14 mg/dl
B = 550 mg/dl
Value
under the
measurement
range.
Value
above the
measurement
range.
If the measured data is below the measurement range, an “@”
mark is added as shown in the example printout.
If the measured data is above the measurement range, a “>” sign
is printed as shown in the example printout.
IMPORTANT
Reported values with “@” marks may not be accurate. Run the
measurement again.
Touch the Lot No. for which you want to enter a lot
compensation coefficient.
NOTE: If is touched, all lot compensation coefficient settings
are printed.