102
NOTE: The following section shows the default.
LAN settings
FDC settings
IP address:192.168.100.1 Subnet mask:255.255.255.0 Port number:65535
PC settings
IP address:192.168.100.2 Port number:65535
NOTE: After configuration, the analyzer shuts down once. When the analyzer is started again, the settings are applied.
8.4.3 Calculated parameter
You can configure the settings to calculate the following items from simultaneously measured values and then display the
results. Items registered for calculation have their measurement results displayed and printed.
Test Name
Calculation
Test Name
Unit Calculation Formula
Globulin GLOB g/dL or g/L GLOB = TP value - ALB value
Albumin/Globulin ratio ALB/GLOB ALB/GLOB = ALB value/(TP value - ALB value)
BUN/Creatinine ratio BUN/CRE mg/mg or (SI)
GOT/GPT ratio (AST/ALT ratio)
GOT/GPT
(AST/ALT)
GOT/GPT = GOT value/GPT value (AST/ALT = AST
value/ALT value)
NOTE: The calculation test name can be changed by
[Test item name setting] in the Mode function for
administrator. Section 8.4.17
Sodium/Potassium ratio Na/K Na/K = Na value/K value
Anion Gap Anion Gap mEq/L or mmol/L Anion Gap = Na value - (Cl value + TCO
2
value)
If using HeskaView Integrated Software® or Heska Data Capture Utility (DCU), desired calculated parameters can be selected
within the software program and do not need to be selected in the analyzer.
IMPORTANT
Calculation items display a value calculated from simultaneously measured values. Therefore, calculation results are
impacted by errors included in measurement values and the effects of interfering materials. Please understand that when
you use the results for a diagnosis, you must make an overall determination in combination with the measurement values
is used in the calculation.
NOTE: In the case of additional testing and retesting, items subject to calculation must be measured at the same time.
NOTE: When there are two or more slides of the same item subject to calculation in the slides measured at the same time, the
initial measurement value is used.
NOTE: When the measurement results of items subject to calculation exceed the measurement range (determination range),
the calculation results of this calculation item are displayed and printed as "****".
NOTE: When an item subject to calculation has expired, an expiration warning symbol (#) is displayed and printed for the
measurement results of this item and the calculation results.