78
7.1.2 Error code and reference section table
Error code Error Description Reference Section
E0021 The disposal box was open during measurement. 7.2.13
E0035 Photometric control system error. 7.2.8
E0050 Light source lamp is burned out. 7.2.5
E0080, E0081 Tip detection sensor error. 7.2.13
E0082–E0085 Tip mounting error. 7.2.4
E0086, E0087 Tip disposal error. 7.2.4
E0111, E0112 Clogging error during sampling of sample. 7.2.4
E0121, E0122 Clogging error during sampling of diluent. 7.2.4
E0124 Clogging error during sampling of diluted sample. 7.2.4
E0125 Clogging error during sampling of diluted sample. 7.2.4
E0127, E0128 Clogging error during sampling of reference fluid. 7.2.4
E0154 Mixing cup error. 7.2.4
E0200–E0204
E0210–E0213
E0220–E0227
Slide transfer error. 7.2.6
E0300, E0301, E0302 Reference white plate measurement error during initial operation. 7.2.8
E0509, E0510 Signal system error inside circuit board. 7.2.8
E0511 Light intensity error. 7.2.5
E0515, E0516, E0517 Signal system error inside circuit board. 7.2.8
E0530, E0531 ISE measurement unit (self-test) error. 7.2.11
E0532, E0533, E0537,
E0538
Signal system error inside circuit board. 7.2.8
E0534, E0535, E0536,
E0539, E0540
Signal system error inside circuit board. 7.2.8
E0565 ISE temperature control error. 7.2.7
E0600-E0604 Incubator motor error. 7.2.6
E0700-E0703 ISE measurement unit motor error. 7.2.6
E0900-E0903 Interference filter motor error. 7.2.5
E1000–E1003,
E1010–E1013
Sampler vertical operation error. 7.2.4
E1100–E1103 Sampler horizontal operation error. 7.2.4
E1200–E1203 Syringe operation error. 7.2.4
E1400–E1403 Sample application unit feed error. 7.2.6
E1500, E1501, E1505 Temperature control error. 7.2.7
E1502, E1503, E1504 Temperature control error. 7.2.7
E1510, E1511, E1512,
E1514, E1515
ISE temperature control error. 7.2.7
E1520, E1521 Installation environment temperature error. 7.2.7
E1530, E1531, E1532 Signal system error inside circuit board. 7.2.8
E1600–E1603 Sample disk operation error. 7.2.6
E1750–E1755 Slide reading control system error. 7.2.8
E3900 Signal system error inside circuit board. 7.2.8
E4401, E4403 Signal system error inside circuit board. 7.2.8
E4410 Measurement result reference error. 7.2.8
E4411 Control chart reference error. 7.2.8
E4454. E4490,
E4600–E4613
Signal system error inside circuit board. 7.2.8
E4730, E4732 Signal system error inside circuit board. 7.2.8
E5101, E5102, E5103 Printer error. 7.2.2