EasyManua.ls Logo

Heska Element DC5X - Cleaning Transfer Bar, Incubator, Spotting Part, and ISE Unit

Heska Element DC5X
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
59
1
2
6.3 Inspecting and Cleaning the Transfer Bar, Incubator, Spotting Part and ISE Unit
Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE unit. These parts should be
checked and cleaned at least once every three months. When a slide transfer error or a tip ejecting error occurs, clean the
spotting part.
6.3.1 Reference plate level check
Perform [Reference plate level check] before cleaning.
1. Touch .
3. Remove dust adhering to the filter with a vacuum cleaner or running
water.
NOTE: If you wash the filter with running water, make sure it is dried well
before setting in the analyzer.
4. Insert the filter into the louver and put the louver back in the analyzer.
Insert the back side of the louver first, then snap the front side in
place.
IMPORTANT
If the analyzer is used without setting the filters, measurement values may
be negatively affected.
2. Touch NORMAL MODE.

Table of Contents

Other manuals for Heska Element DC5X

Related product manuals