2
WARNING
When discarding the main unit that may be contaminated with blood samples, follow applicable regulations in your
country and dispose of it appropriately.
WARNING
When handling samples (blood) and performing maintenance (cleaning the analyzer), always follow biohazard procedures
(wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If the sample is
accidentally touched, immediately rinse the contaminated area thoroughly under running water and disinfect.
1.4 Warnings Regarding Explosives
WARNING
Do not use flammable and explosive gas around the equipment.
1.5 Warnings Regarding Electric Shock
WARNING
The voltage supplied to the equipment is AC 100–240 V. To avoid electrical shock, observe the following precautions.
Avoid installation locations in which water may splash on the equipment.
Plug the power cable into an outlet with a grounding receptacle. Electrical shock may occur if the equipment is not
grounded to a protective earth.
Make sure that all cables have been properly connected.
WARNING
Do not remove covers or other parts that are secured with screws to avoid electrical shock that may result from exposure
to hazardous voltage or injury from moving parts.
1.6 Electromagnetic Compatibility (EMC)
NOTE: It is the manufacturer's responsibility to provide equipment electromagnetic compatibility information to the customer
or user. On the other hand, it is the users responsibility to ensure that a compatible electromagnetic environment for the
equipment can be maintained so the device will perform as intended.
This equipment conforms to the following emission and immunity requirements:
IEC 6132626:2012 (Class B)
EN 6132626:2013 (Class B)
IEC 6132626:2005 (Class B)
EN 6132626:2006 (Class B)
This analyzer has been designed and tested to CISPR 11 Class B. In a domestic environment it may cause radio interference, in
which case, you may need to take measures to mitigate the interference.
The electromagnetic environment should be evaluated prior to operation of the device.
Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g., unshieided intentional RF
sources), as these may interfere with the proper operation.
If this equipment does cause harmful interference to other devices, which can be determined by tuning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures.
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.