Heska Element DC5X - Replacing the Sampler O-Ring

Heska Element DC5X
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72
If the check results are [NG], touch REPLACE ORING, then
replace the O-ring according to the following procedure.
NOTE: The check results can be printed by touching .
7. Once the check ends, open the sampler cover, and then
remove the sampler leak check tool mounted on the sampler
nozzle by hand.
6.6.2 Replacing the sampler O-ring
The sampler nozzle O-ring wears over time. Replace it if indicated.
IMPORTANT
If you do not replace the O-ring when indicated, the sample application volume will be inaccurate, which may negatively
affect the measurement values.
IMPORTANT
Be sure to use the O-ring designed for the Element DC5X Analyzer. Otherwise, the sample application volume will be
inaccurate, and the measurement results may be negatively affected.
1. Remove the sample rack and slide cartridge.
2. Touch .