Error code Error Description Reference Section
W4720
Clock time error. 7.2.8
W4721, W4731,
W4733, W4735,
W4736
Signal system error inside circuit board. 7.2.8
W5000, W5020
Signal system error inside circuit board. 7.2.8
W5050
USB memory has not been inserted. 7.2.8
W5100
Printer error. 7.2.2
W5110
No recording paper. 7.2.2
W5330, W5331
Sample barcode reader error. .2.12
W5400
Expired slide. 7.2.10
W5401, W5402
Tip discarding error. 7.2.4
W66xx
Input value warning. 7.2.13
I5920
Reboot warning. 7.2.13
7.1.3 Printout reference table
Printout Meaning and Conditions Troubleshooting
H
Measurement value is higher than the upper limit of the preset
reference interval.
L
Measurement value is lower than the lower limit of the preset
reference interval.
>
Measurement value is higher the upper limit of the measurement
range.
<
Measurement value is lower than the lower limit of the
measurement range.
@
The measurement values are reference values outside the
measurement range.
IMPORTANT
Measurement values might be incorrect.
+ OR
Measurement value of the ISE test is higher than the upper limit of
the measurement range.
Refer to Section 7.2.11
– OR
Measurement value of the ISE test is lower than the lower limit of
the measurement range.
Refer to Section 7.2.11
+, –,$
IMPORTANT
Measurement values might be incorrect. Refer to the related
troubleshooting pages and rerun the measurement.
Refer to Section 7.2.7
*, ? Light intensity is fluctuating during measurement. Refer to Section 7.2.5
E
Unable to calculate the measurement result because measurement
was interrupted due to a transfer problem or other error.
The measurement results are printed as [****].
Check whether other errors have occurred, and then perform the
necessary troubleshooting.
&
Abnormally high measurement value. In the case of a Ca-P slide, the
sample has not been spotted.
Perform a dilution.
#
The valid term of the slide has expired.
Measurement value may not be accurate.
IMPORTANT
The # for expired slides is not shown on ISE tests.
Use slides within their validity terms.
¥ Problem with sample application on slide (not applied). Rerun the measurement.