55
0.5
0.3
0.1
FUJI PLAIN TUBE
(1.5 ml)
Pour reference fluid
within the range.
FUJI PLAIN TUBE
(0.5 ml)
0.5
1.0
5.5 Loading Reference Fluid
Reference fluid is necessary for ISE measurements. Section 3.5
WARNING
When the analyzer is applying samples, touch PAUSE to interrupt sample application (Section 4.1), and then load the
reference fluid. It is unsafe to set tips during sample application due to the moving parts of the sampler and the sample
disk.
IMPORTANT
Use a DRI-CHEM PLAIN TUBE (0.5 or 1.5 mL) for reference fluid. Otherwise, measurement errors or malfunctions may occur.
Configure the size of tube to be used via [Tube setting for diluent and reference fluid] in Mode function for administrator.
Section 8.4.12
IMPORTANT
Replace DRI-CHEM PLAIN TUBES (0.5 mL and 1.5 mL) every 12 hours.
1. Prepare reference fluid.
Pour the following amounts into each type of FUJI PLAIN TUBE.
For 1.5 mL tube: 1.0-1.5 mL
For 0.5 mL tube: 0.3-0.5 mL
NOTE: Use reference fluid after it reaches room temperature.
2. Make sure [Warming up] or [Ready to test] displayed on the LCD.
Make sure the analyzer is not applying samples.
3. Touch reference fluid display on the LCD.