18
3.2 Measurement Preparation
Cleaning the disposal box
1. Remove the disposal box and dispose of the tips and slides.
2. Use isopropyl alcohol to disinfect the waste drawer.
IMPORTANT
If the disposal box is full, a slide transfer error or tip disposal error may
occur, which may negatively affect the measurement values.
IMPORTANT
After emptying the disposal box, be sure to load the disposal box. If
the analyzer is used without the box, the measurement results may be
negatively affected.
Checking the recording paper
1. Press the area shown in the illustration to release the lock (1), and
then lift up the cover (2).
2. Check the remaining recording paper (FUJI thermal paper). If the
recording paper is short for the measurement, replace the old roll
with a new one.
NOTE: Before closing the cover, pull out the edge of the recording paper.
2
1
Cover
Press the
concave
section.
paper
Recording
Cutter
opening
Recording
paper