60
In the case of Level 1
soiling (0-49%)
In the case of Level 2
soiling (50-79%)
In the case of Level 3
soiling (80% or higher)
3. Touch REFERENCE PLATE LEVEL CHECK to run a check.
If check ends, the Check results screen appears.
4. Check the results of [Reference plate level check].
If the result is level 2 or level 3, cleaning is necessary. Touch OK. Using the procedure on the following pages, clean the
transfer bar, incubator, spotting part, and ISE unit.