Operation
When using external devices in the interior, the
Switchboard option is recommended for reasons of
operational safety.
NOTICE
When operating a thermal load in the interior, a suit-
able signal transducer for monitoring the incubator
must be connected to the potential-fr
ee alarm con-
tact.
Fig. 206: Heat compensation at an ambient temperature of 22 °C
T [°C] Temperature in °C
P [W] Power in Watts
1 Lowest reachable temperature, HettCube 600 R (0 °C to +65
°C)
2 Lowest reachable temperature, HettCube 400 R (0 °C to +65
°C)
3 Lowest reachable temperature, HettCube 200 R (0 °C to +65
°C)
The lowest reachable temperature values when using external
devices with different heat outputs in the interior of the incubator.
The data refers to incubators with standard equip-
ment.