en
12.2020 / Rev. 02
AB66005
124 / 174
Cleaning, disinfection and maintenance
Fig. 213: Maintenance
Recommendation
7.4 Autoclaving
Autoclaving accelerates the ageing process of plas-
tics. In addition, it can cause discolourations in plas-
tics.
The slide-in modules, drawers, support rails, telescope rails, latching
rails and plates in the interior can be autoclaved at 121°C / 250°F
(20 Min).
Before autoclaving, these parts must be removed from the interior.
Nothing denitive can be said about the degree of sterility.
7.5 Maintenance
WARNING
There is a risk that the sample temperature control will
be faulty
.
There is a risk that the temperature control of micro-
biological samples is not correct, which could result in
a misinterpretation of the test results or the destruc-
tion of the samples.
Have maintenance and calibration done.
In order to ensure the longevity and error-free running of the incu-
bator
, the incubator should be maintained at regular intervals by the
manufacturer or personnel authorized by the manufacturer.
The manufacturer offers maintenance contracts for this purpose. The
operating company is obligated to make sure that the maintenance is
carried out by an authorized service technician.
Interval Maintenance work Personnel
Before every use Before every use, make sure that the incubator is in
perfect technical condition and no r
estrictions are to
be expected according to the desired application.
User
Annual maintenance Check by a service technician from the manufacturer. Service technician
Check by a service technician from the manufacturer.
The information can be closed with the
[Close]
button.
Service technician
After 2 years, a message appears automatically with a notice about
the incubator being maintained by a service technician from the man-
ufacturer. The user can make the message disappear; afterwards,
the message appears again at yearly intervals. Once maintenance is
performed by the service technician, the time period until the mes-
sage appears again is set to 2 years.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals