Fig. 84: ‘Manual mode
Set
temperature
Keypad entry’
1
[Clear input]
button
2 Keypad
2.
Fig. 85: ‘Manual mode
Start’
1
[Start manual mode]
button
Fig. 86: ‘Manual mode
Start
Main
screen’
Fig. 87: ‘Manual mode
Start
Start
delay’
You can now either
select the existing value by tapping it and entering a new
value via the keypad or
delete the existing value with the
[Delete entry]
button and
entering a new value via the keypad.
Conrm the entry with the
[Conrmation]
button.
6.11.2 Starting manual mode
T
o start manual mode, proceed as follows:
1. Make all settings for manual mode
Chapter 6.11.1 ‘Manual
mode settings’ on page 54
.
2. Start manual mode with the
[Start manual mode]
button.
NOTICE
A temperature guard is always activated and
cannot be switched of
f. If the temperature set-
point lies outside of the temperature range of
set temperature guards, an error message is
displayed when manual mode is started.
If no start delay for manual mode was selected, you will go
to the main scr
een.
3. If a start delay was specied, a message indicating when
manual mode will start will appear befor
ehand.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals