Cleaning, disinfection and maintenance
Surface cleaning and care
Slide-in modules and drawers in the
interior
Fig. 207: Removing the slide-in module
In the event of strong contamination, the lter mat
must be exchanged. A service technician must do
the exchanging.
For incubators with cooling, we r
ecommend that the
lter mat be exchanged once a year.
7.1 Cleaning
WARNING
The incubator might be contaminated.
There is a risk of contamination by bacteria, viruses or
other pathogens. Based on its own hazard analysis,
the operating company must dene whether and how
personal protective equipment is to be worn.
■ Clean the housing and the interior of the incubator regularly
and, if necessary
, clean with soap or a mild detergent and a
moist cloth. This is for hygienic purposes and also prevents
corrosion due to adhering contaminants.
■ Ingredients of suitable detergents: soap, anionic tensides,
non-ionic tensides.
■ After using cleaning agents, remove any cleaning agent res-
idue by wiping with a damp cloth.
■ The surfaces must be dried immediately after cleaning.
■ The interior is to be checked for damage once a year.
For easier cleaning of the interior
, the slide-in modules and drawers
can be removed from the interior.
Removal:
1. Remove the slide-in modules and drawers.