EasyManuals Logo

Hettich HettCube 200 User Manual

Hettich HettCube 200
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
137 / 174
en
12.2020 / Rev. 02
AB66005
Technical data
EMC
Emitted interference, immunity to
interfer
ence
EN/IEC 61326-2-6, Class B
Type of protection (EN 60529) IP 20
Temperatures
Temperature range 1 K above the ambient temperature up to 65 °C
Temperature setting precision 0.1°C
Temperature deviation over time at
37 °C
± 0.1 K
Spatial temperature deviation at
37 °C
± 0.2 K
Spatial temperature deviation at
25 °C
± 0.1 K
Recovery time (after door open for
30 s) at temperatur
e setpoint 37
°C
4.5 min
Dimensions
Interior dimensions
W x D x H 535 x 690 x 850 mm
Exterior dimensions
W x D
1)
x H 710 x 825
1
x 1425 mm
Weight approx. 117 kg
3)
Maximum load per standard feed 50 kg
Maximum total load 100 kg
1)
W
ithout door handle and access port (Ø 42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +10 kg.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.4 HettCube 400 R
Manufacturer Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model HettCube 400 R
Type 64005 64005-01

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals