Fig. 214: Activating the automatic
circuit breaker
8 Troubleshooting
If the error cannot be remedied using the trouble-
shooting table, notify the service technician. Please
specify the type of device and the serial number.
Both numbers can be found on the name plate of
the incubator.
If a malfunction occurs, there is an optical and
acoustic alarm.
After a malfunction (power failure, door open too
long, etc.), the operating company must assess
whether the samples are still usable. He must there-
fore have the corresponding qualication.
8.1 Activating the automatic circuit breaker
For the Switchboard option (external socket strip)
only.
1. Switch off the mains switch and disconnect the incubator from
the mains!
2. Press the plastic pin (1) of the automatic circuit breaker.
3. Reconnect the incubator to the mains.