Technical data
Spatial temperature deviation at
25 °C
± 0.1 K
Recovery time (after door open for
30 s) at temperatur
e setpoint 37
°C
3 min
Dimensions
Interior dimensions
W x D x H 535 x 690 x 420 mm
Exterior dimensions
W x D
1)
x H 710 x 825
1
x 970 mm
Weight approx. 92 kg
3)
Maximum load per standard feed 50 kg
Maximum total load 80 kg
1)
W
ithout door handle and access port (Ø 42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +6 kg.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.2 HettCube 200 R
Manufacturer Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model HettCube 200 R
Type 62005 62005-01
Mains voltage (± 10 %) 220-240 V 1~ 100-120 V 1~
Mains frequency 50-60 Hz
Connected load 450 VA/850 VA
2)
500 VA/900 VA
2)
Energy consumption at 37°C 0.038 kWh/h
Cooling medium R 513 a
(Contains uorinated gr
eenhouse gases documented by the Kyoto Pro-
tocol)
Amount of refrigerant 200 g
Global-warming potential (GWP) 631
Inner volume 150 l

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals