Technical data
Ambient conditions (EN / IEC 61010-1)
Set-up site Indoors only
Height Up to 2000 m above sea level
Ambient temperature 16°C to 35°C
Humidity Maximum relative humidity 75%, non-condensing.
Overvoltage category (IEC
60364-4-443)
II
Degree of contamination 2
Device protection class I
Noise level 44 dB(A)
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference, immunity to
interfer
ence
EN/IEC 61326-2-6, Class B
Type of protection (EN 60529) IP 20
Temperatures
Temperature range 0 °C to 65 °C
4)
Temperature setting precision 0.1°C
Temperature deviation over time at
37 °C
± 0.1 K
Spatial temperature deviation at
37 °C
± 0.2 K
Spatial temperature deviation at
25 °C
± 0.1 K
Recovery time (after door open for
30 s) at temperatur
e setpoint 37
°C
3 min
Dimensions
Interior dimensions
W x D x H 535 x 690 x 420 mm
Exterior dimensions
W x D
1)
x H 710 x 825
1
x 970 mm
Weight approx. 103 kg
3)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals