EasyManuals Logo

Hettich HettCube 200 User Manual

Hettich HettCube 200
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
en
12.2020 / Rev. 02
AB66005
162 / 174
Examples
Fig. 225: ‘Dehumidication’ program
B.5 Temperature drop on the weekend including the passive dehumidication
option
Description:
The incubator is operated at a temperatur
e of 37 °C during the work-
week. Starting on Saturday 12:00 p.m., the temperature drops to 4
°C. On the following Monday, the sample material is taken out at
approx. 8:00 a.m. at a temperature of 37 °C. The beginning of the
heating phase is dened with a lead time of 5 h on Monday at 3:00
a.m. During these 5 h, the ventilation opening should be open. This
way, moisture can escape from the incubator during the heat-up
phase.
Operating mode:
Program mode, programming with external contact.
Mon
P
S1
PS2PS2
PS3
PS4PS4
Tue Thu Fri Sat Son MonWed
37
°C
4
1
PS3
2
2
Fig. 224: Diagram ‘Using the passive dehumidication option,
example of temperature drop during weekend’
1 Starting time (Friday 10:00 a.m.)
2 Ventilation opening open
PS Program segment
NOTICE
The device must be equipped with the passive dehu-
midication option
Chapter 6.13.6.2 ‘Passive
dehumidication option’ on page 104
.
Set the program as shown; see
Chapter 6.12.2
‘Creating a program’ on page 70. The ventilation
opening is opened as long as the control contact is
active.
Please take the starting time into account; see
‘Timer’ on page 86.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals