Fig. 50: Screw for separate blocking of
control panel
1 Screw
Fig. 51: Control panel blocking
1 Bore with thread,
[Control panel
blocking]
6.4 Fixing the mechanical display protection of the control panel
To prevent the incubator from being operated by unauthorized per-
sonnel, the mechanical display protection can be additionally
blocked.
1.
In the delivery condition, there is a screw on
the rear side of the incubator, which can be
used to block the mechanical display protec-
tion.
Remove this screw.
2. Push the mechanical display protection to the right.
3. Screw the screw into the
[Control panel blocking]
bore.
The mechanical display protection is blocked. The door can
still be opened and closed.
6.5 Standard check before every use
Every time befor
e using the incubator, make sure that
■ The mains voltage agrees with the specication on the name
plate.
■ The incubator is standing at on its feet.
■ There is at least one slide-in module in the incubator.
■ The door closes tightly.
■ No alarm or error messages are pending.
■ The ventilation slots of the incubator are free.
■ The incubator has no obvious damage.