Technical data
Spatial temperature deviation at
25 °C
± 0.1 K
Recovery time (after door open for
30 s) at temperatur
e setpoint 37
°C
4.5 min
Dimensions
Interior dimensions
W x D x H 535 x 690 x 850 mm
Exterior dimensions
W x D
1)
x H 710 x 825
1
x 1425 mm
Weight approx. 128 kg
3)
Maximum load per standard feed 50 kg
Maximum total load 100 kg
1)
W
ithout door handle and access port (Ø 42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +10 kg.
4)
Lowest settable temperature -5 °C. Reaching a temperature < 0°C
depends on the ambient conditions, however.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.5 HettCube 600
Manufacturer Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model HettCube 600
Type 66000 66000-01
Mains voltage (± 10 %) 220-240 V 1~ 100-120 V 1~
Mains frequency 50-60 Hz
Connected load 450 VA/850 VA
2)
500 VA/900 VA
2)
Energy consumption at 37°C 0.056 kWh/h
Inner volume 520 l
Ambient conditions (EN / IEC 61010-1)
Set-up site Indoors only
Height Up to 2000 m above sea level

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals