6 Operation
DANGER
A danger is posed by penetrating liquids.
There is an electric shock hazard for personnel.
− If liquid has penetrated the incubator during oper-
ation, do not touch the incubator and switch it of
f
on the building side (e.g. via the fuse).
− In such a case, contact a service technician of the
manufactur
er.
CAUTION
There is a danger due to vibrations.
A danger is posed by unsecur
ed, vibrating devices on
the slide-in modules.
− During the operation of a vibrating device (e.g., a
shaker), the slide-in module on which the vibrating
device is standing must be secur
ed against falling
out. Hettich recommends the drawer with the
telescopic rail (60024) in combination with the
xing kit (60919).
− When operating a vibrating device (e.g., a shaker),
the operating company must make sur
e that this
cannot move away on the slide-in module. Hettich
accepts no responsibility for damage caused by
the use of external devices.
− When operating a vibrating device, the incubator
must be monitor
ed. It might be necessary to
restrict the speed range of the shaker, for
example.