1
Fig. 212: Right bafe sheet
1 Right bafe sheet
7. Disengage the right bafe sheet upwar
d and then remove
toward the front.
Installation is done in the opposite order.
When installing the upper bafe sheet, push this upward and fasten
with the two knurled screws. It must be ush with the ceiling of the
interior.
7.2 Disinfection
If infectious material gets into the interior, disinfect it immedi-
ately
.
Ingredients of suitable disinfectants: Ethanol, n-propanol,
isopr
opyl alcohol, ethyl hexanol, corrosion inhibitors.
After using disinfectants, remove the disinfectant residue by
wiping with a damp cloth.
The surfaces must be dried immediately after disinfecting.
WARNING
Risk due to insufcient disinfection after contamina-
tion!
Ther
e is a risk of infection for the user. Observe your
laboratory rules (hygiene plan, etc.) for handling bio-
logical agents.
7.3 Removal of radioactive contaminants
The agent must be specically labelled as being an agent
for r
emoving radioactive contaminants.
Ingredients of suitable agents for removing radioactive con-
taminants: anionic tensides, non-ionic tensides.
After removing the radioactive contaminants, remove the
agent r
esidue by wiping with a damp cloth.
The surfaces must be dried directly after removing the radi-
oactive contaminants.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals