Fig. 13: Passive dehumidication
ventilation opening
1 Ventilation opening for passive
dehumidication
Accessories
Icing can occur in the case of continuous cooling
temperatures between +1°C and +5°C. Using pro-
gram mode, the customer can create a suitable
"defrosting" program. The device runs through a
daily cycle in which the cooling unit is briey heated,
and then specically cooled again afterwards. This
prevents ice formation.
3.2.6 Passive dehumidication
The device is equipped with a ventilation opening, which allows
moisture to be released from the incubator. This ventilation opening
can be opened and closed individually or time-controlled via the
touchscreen. It is operated via the control contact function.
This option can specically be used to
■ prevent condensation from forming inside the incubator during a
heating operation fr
om cooling mode.
■ accelerate the drying process of the interior.
NOTICE
Dehumidication is only possible if the moistur
e con-
tent of the air outside of the incubator is lower than
that inside the incubator.
The ambient temperature should be at least 5°C
lower than the temperature to which the incubator is
to be heated.
The passive dehumidication option and Switchboar
d option are
mutually exclusive. Therefore, it is only possible to realize either the
passive dehumidication option
Chapter 6.13.6.2 ‘Passive dehu-
midication option’ on page 104
or the Switchboard option
Chapter 6.13.6.1 ‘Switchboard option’ on page 102
on one incu-
bator.
3.3 Accessories
Brief description of the accessories: Flexible equipment articles which
the device can be equipped with later
Further information about the accessories can be found in the
appendix
Appendix C ‘Accessory overview’ on page 167
.
Shelf (set) 60001
Shelf (HTS set) 60031