Fig. 126: ‘Program mode
Start
settings’
6.12.5 Program mode — Start settings
Before starting program mode, you can change the
start settings.
Item Designation Description/function
1
[Start]
button The currently selected program is started.
2 Program behaviour The program behaviour indicates whether the times of the
program segments start once the temperature of the pro-
gram segment has been reached or after the specied
time.
At temp:
The start time of a program segment does not begin until
the temperature setpoint of this program segment has
been reached.
At start:
The dened time periods of the program segments are run
through in succession without caring whether the tempera-
ture setpoint has been reached.
3
PIN protection
‘PIN protection’
on page 85
To protect the program from being changed, you can acti-
vate/deactivate the PIN protection here.
4
[Timer]
switch The start date and start time of the program can be entered
here.