Operation
Fig. 173: ‘Failure alarm’ menu
Fig. 174: ‘Export’ menu
Fig. 175: Refreshing the USB stick
1
[Refresh]
button
Fig. 176: USB stick export
1
[Export]
button
2 Partition on USB stick
6.13.12 Failure alarm
Y
ou can congure the alarm sound with this setting. The volume, fre-
quency and duration are adjustable. The volume of the failure alarm
can be adjusted in stages.
Via the input
‘Number of reproduced signal sounds’
, it can be speci-
ed how often a signal sound should repeat itself (1 to 10000 and
innitely).
6.13.13 Export
You can export programs, temperature drops, user
instructions and log les to a USB stick via the
export function.
The exported programs and temperature drops can
be transferred directly to other HettCube incubators.
NOTICE
If the USB stick is plugged in upside-down, this can
r
esult in the incubator being switched off.
To export data, proceed as follows:
1. Stick a USB stick into the service interface (USB port) on the
fr
ont side of the incubator.
2. Tap on the
[Refresh]
button.
The partition on the USB stick is displayed.
If there are multiple partitions on your USB
stick, you can select the partition to which
data should be exported via ‘Partition on
USB stick’.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals