Examples
Fig. 231: ‘Cleaning mode’ program
B.8 Incubator is always cleaned on Wednesdays due to the laboratory routine
(SOP) (standstill mode)
37
°C
4
PS1
PS2
P
S1
1
Mon Tue Thu Fri Sat SunW
ed Mon
Fig. 230: Diagram ‘Incubator is always cleaned on Wednesdays due
to the laboratory routine (SOP) (standstill mode)’
1 Starting time (Thursday, 12:00 a.m.)
PS Program segment
Operating mode:
Program mode; Always an incubation temperature and one-time
standstill segment in every week
Set the program as shown
Chapter 6.12.2 ‘Cre-
ating a program’ on page 70.
During standstill mode, the incubator is not regu-
lated, but the current interior temperature is dis-
played.
Please take the starting time into account
‘Timer’
on page 86.