EasyManuals Logo

Hettich HettCube 200 User Manual

Hettich HettCube 200
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #165 background imageLoading...
Page #165 background image
165 / 174
en
12.2020 / Rev. 02
AB66005
Examples
Fig. 231: ‘Cleaning mode’ program
B.8 Incubator is always cleaned on Wednesdays due to the laboratory routine
(SOP) (standstill mode)
37
°C
4
PS1
PS2
P
S1
1
Mon Tue Thu Fri Sat SunW
ed Mon
Fig. 230: Diagram ‘Incubator is always cleaned on Wednesdays due
to the laboratory routine (SOP) (standstill mode)’
1 Starting time (Thursday, 12:00 a.m.)
PS Program segment
Operating mode:
Program mode; Always an incubation temperature and one-time
standstill segment in every week
Set the program as shown
Chapter 6.12.2 ‘Cre-
ating a program’ on page 70.
During standstill mode, the incubator is not regu-
lated, but the current interior temperature is dis-
played.
Please take the starting time into account
‘Timer’
on page 86.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals