EasyManua.ls Logo

Hettich HettCube 200 - Language; Sound

Hettich HettCube 200
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
105 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Fig. 164: ‘Language’ menu
Fig. 165: ‘Sound’ menu
1
[Buttons]
controller
2
[Program end]
controller
3 Input of
‘Number of reproduced
signal sounds’
NOTICE
If the ventilation opening is opened during cooling
mode, ther
e is a danger of moisture from the envi-
ronment getting into the incubator.
At target temperatures below 10 °C, the ventila-
tion opening must be kept closed. Otherwise,
ther
e is a danger the evaporator will ice up.
If the ventilation opening is open
the spatial temperature deviation will increase
by ± 0.1 °C as compar
ed to the value speci-
ed in the technical data.
the current consumption specied in the tech-
nical data will incr
ease, depending on the
ambient temperature, by approx. 0.016 kW/h.
In heating mode, when the ventilation opening is
open continuously
, there is a risk of the biological
sample material drying out.
6.13.7 Language
In this menu, you can select the display language.
The display language is set during commissioning. If
another language is desired for the displays, these
can be selected from a list. The displayed language
changes directly after selecting a new language set-
ting.
To select the desired language, tap it.
6.13.8 Sound
The volume for the keypad feedback can be set in 4 levels or muted
with the
[Buttons]
controller.
The volume can be set in 4 levels or switched off for the signal sound
after the end of the program via the
[Program end]
controller.
Via the input
‘Number of reproduced signal sounds’
, it can be speci-
ed how often a signal sound should repeat itself (1 to 10000 and
innitely).

Table of Contents

Related product manuals