EasyManuals Logo

Hettich HettCube 200 User Manual

Hettich HettCube 200
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
31 / 174
en
12.2020 / Rev. 02
AB66005
Commissioning
1
Fig. 24: Levelling the incubator
1 Adjustable device foot
Power supply connection
1. Set up and level the incubator stably on a at, non-ammable
surface.
For HettCube 400/400 R and HettCube 600/600 R incubators
only:
Screw the levelling elements (A) that are screwed into the
device feet far enough downwar
d that they touch the oor
and the rollers are completely relieved.
Align the incubator horizontally by turning the levelling ele-
ments.
Screw the adjusting foot on the door downward until it is
appr
ox. 7 mm above the oor to keep the incubator from
tipping. Screw the hexagon nuts upward and tighten to
secure the adjusting foot.
For HettCube 200/200 R incubator only:
Align the incubator horizontally by turning the levelling ele-
ments scr
ewed into the device feet.
For incubator with glass door only: Screw the adjusting foot
on the door downwar
d until it is approx. 7 mm above the
oor to keep the incubator from tipping. Screw the hexagon
nuts upward and tighten to secure the adjusting foot.
Only for devices of size 400 and 600: Transport securing
device
CAUTION
Two incubators may only be stacked on top of
each other in the case of HettCube 200/200 R.
The upper incubator must be xed to the lower
incubator with the stack kit (or
der no. 60009)
and additionally secured against tilting. We rec-
ommend the xing kit (order no. 60012) to
securely fasten the upper incubator to the wall.
2. If required, adjust the heights of the slide-in modules and
drawers; see
Chapter 5.3.1 ‘Standard slide-in modules’
on page 38
and
Chapter 5.3.2 ‘Telescopic slide-in modu-
les’ on page 39
.
5.2.2 Incubator connection
DANGER
A hazard is posed by a damaged mains supply line.
Ther
e is an electric shock hazard for personnel.
Check the incubator and the mains supply line for
damage. If damage is discover
ed, the incubator
must not be operated.
The incubator may only be operated at a socket
with a personal pr
otection switch.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals