Fig. 24: Levelling the incubator
1 Adjustable device foot
Power supply connection
1. Set up and level the incubator stably on a at, non-ammable
surface.
For HettCube 400/400 R and HettCube 600/600 R incubators
only:
■ Screw the levelling elements (A) that are screwed into the
device feet far enough downwar
d that they touch the oor
and the rollers are completely relieved.
■ Align the incubator horizontally by turning the levelling ele-
ments.
■ Screw the adjusting foot on the door downward until it is
appr
ox. 7 mm above the oor to keep the incubator from
tipping. Screw the hexagon nuts upward and tighten to
secure the adjusting foot.
For HettCube 200/200 R incubator only:
■ Align the incubator horizontally by turning the levelling ele-
ments scr
ewed into the device feet.
■ For incubator with glass door only: Screw the adjusting foot
on the door downwar
d until it is approx. 7 mm above the
oor to keep the incubator from tipping. Screw the hexagon
nuts upward and tighten to secure the adjusting foot.
■ Only for devices of size 400 and 600: Transport securing
device
CAUTION
Two incubators may only be stacked on top of
each other in the case of HettCube 200/200 R.
The upper incubator must be xed to the lower
incubator with the stack kit (or
der no. 60009)
and additionally secured against tilting. We rec-
ommend the xing kit (order no. 60012) to
securely fasten the upper incubator to the wall.
2. If required, adjust the heights of the slide-in modules and
drawers; see
Chapter 5.3.1 ‘Standard slide-in modules’
on page 38
and
Chapter 5.3.2 ‘Telescopic slide-in modu-
les’ on page 39
.
5.2.2 Incubator connection
DANGER
A hazard is posed by a damaged mains supply line.
Ther
e is an electric shock hazard for personnel.
− Check the incubator and the mains supply line for
damage. If damage is discover
ed, the incubator
must not be operated.
− The incubator may only be operated at a socket
with a personal pr
otection switch.