6.8.3 Functional description of program mode
V
ia program mode, you can combine complex sequences of temper-
atures and durations to form one program.
A program is composed of program segments. A program segment
can also be dened as a standstill segment (incubator in standstill
without regulation).
Each program segment consists of a temperature and a duration.
The duration is calculated via the start time and end time or the end
time is calculated from the start time and duration.
It can be read off from the rotating circular ring in which program
segment the program is currently located. Firstly, the number of the
currently run-through program segment is displayed in the middle of
the circular ring. Secondly, the circular ring itself is subdivided into as
many parts as there are program segments. The number of circular
ring segments shown in black corresponds to the number of the pro-
gram segment that is currently being run.
Example:
The program consists of 4 segments.
When the rst segment is being run thr
ough, the
length of the black lled-out segment of the cir-
cular ring is ¼ of the total circle.
When the second segment is being run through,
the length is ½ of the total cir
cle, etc.
When the last segment is being run through, the
r
otating circular ring is lled out completely in
black, as shown on the screen above.
In the case of cooled incubators, each program segment can addi-
tionally control the control contact (On/Off)
Chapter 6.13.6 ‘Con-
trol contact (option)’ on page 101
.
By saving the program, only the temperatures and durations of the
program segments are saved. During the planning, a start date and
start time can be dened. This way, a daily routine or a specication
can be simply programmed.
The following additional functions can be activated by starting a pro-
gram:
Program
‘at temp.’
, program
‘at start.’
After the program runs, specify a hold temperature.
Activate the control contact (for cooled incubators).
Time-delayed start of program mode.
Repeat program.
If the incubator is switched off while manual or pro-
gram mode is running, when the incubator is
switched on, it indicates that a power failure was
registered.
If program repetitions were dened in the start settings, these are
executed
‘Program repetition’ on page 86
.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich HettCube 200 and is the answer not in the manual?

Hettich HettCube 200 Specifications

General IconGeneral
BrandHettich
ModelHettCube 200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals