Mains voltage (± 10 %) 220-240 V 1~ 100-120 V 1~
Mains frequency 50-60 Hz
Connected load 450 VA/850 VA
2)
500 VA/900 VA
2)
Energy consumption at 37°C 0.046 kWh/h
Cooling medium R 513 a
(Contains uorinated gr
eenhouse gases documented by the Kyoto Pro-
tocol)
Amount of refrigerant 200 g
Global-warming potential (GWP) 631
Inner volume 150 l
Ambient conditions (EN / IEC 61010-1)
Set-up site Indoors only
Height Up to 2000 m above sea level
Ambient temperature 16°C to 35°C
Humidity Maximum relative humidity 75%, non-condensing.
Overvoltage category (IEC
60364-4-443)
II
Degree of contamination 2
Device protection class I
Noise level ≤ 44 dB(A)
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference, immunity to
interfer
ence
EN/IEC 61326-2-6, Class B
Type of protection (EN 60529) IP 20
Temperatures
Temperature range 0 °C to 65 °C
4)
Temperature setting precision 0.1°C
Temperature deviation over time at
37 °C
± 0.1 K
Spatial temperature deviation at
37 °C
± 0.2 K