No. Fault description Cause Remedy
37 Error 37 The temperature set in temper-
atur
e guard 3.2 was exceeded.
The set temperature setpoint is
higher than the set temperature
of class 3.2.
Please acknowledge the message
with
[X]
; The customer decides
what to do next after checking the
samples. The temperature guard
settings must be adapted
Chapter 6.13.5 ‘Temperature
guard’ on page 98
.
The temperature set in temper-
ature guard 3.2 was fallen short
of. The set temperature setpoint
is lower than the set tempera-
ture of class 3.2.
Please acknowledge the message
with
[X]
; The customer decides
what to do next after checking the
samples. The temperature guard
settings must be adapted
Chapter 6.13.5 ‘Temperature
guard’ on page 98
.
The temperature set in temper-
atur
e guard 3.2 was fallen short
of. The door was open for too
long and the actual temperature
deviates from the temperature
setpoint.
38 Error 38 The door is open. Please close the door.
39 Error 39 Door was closed. Message
‘No 38’
remains for three
seconds and then disappears. The
message can be more quickly
removed by acknowledging with
[X]
.
40 Error 40 The current program is
stopped.
If the
[Start]
button is pressed, the
program continues.
42 Error 42 This is not an error, but a
warning for the user that there
was a power failure. The pro-
gram is continued based on the
user settings (Power failure/
Time).
Please acknowledge the message
with
[X]
; The customer decides
what to do next after checking the
samples.
44 Error 44 No SD card was detected. Restart the system. Check whether
further operation is possible.
If the err
or appears again: Call
Service; Provide the device type,
serial number and error number.
45 Error 45 The communication between
the touchscr
een and controller
is interrupted.
Restart the system. Check whether
further operation is possible.
If the error appears again: Call
Service; Provide the device type,
serial number and error number.