■ The drying rate of the culture medium during incubation is inu-
enced by various factors, such as
– ambient conditions (e.g., relative humidity of the environment)
– handling and working method (e.g., previous processes, dura-
tion, and fr
equency of doors opening)
– culture or incubation conditions (e.g., incubation time and
temperatur
e).
These inuences should be especially considered in the case of
longer incubation times.
■ The sample material should not be placed outside of the dened
utilized space
Chapter 9.7 ‘Denition of the utilized space’
on page 143
. The specied temperature data refer to the dened
utilized space.
■ Repairs may only be performed by a person authorized by the
manufactur
er.
■ Only original spare parts and approved original accessories of the
company Andr
eas Hettich GmbH & Co. KG may be used.
■ The safety and reliability of the incubator are only guaranteed if:
– the incubator is operated according to the operating instruc-
tions.
– the electrical installation at the set-up site of the incubator
meets the r
equirements of the EN/IEC regulations.
The incubator is safe if used as intended and
observing the descriptions and information given in
this documentation.