138
Data That Can Be Saved and Loaded
IMPORTANT
Hioki guarantees the operation of our optional SD Memory Card and USB Drive only. Operation
of any other storage media cannot be guaranteed.
Preparing for power outages and conguring associated settings
NOTICE
Do not use damaged media.
Doing so could prevent processing to close les from completing in the anticipated
amount of time, corrupting les.
Measurement data will be lost if power is interrupted during measurement. It is recommended to
make the following preparations and settings when measuring data for an extended period of time.
• Use the Z1007 Battery Pack.
When using the optional Z1007 Battery Pack, the instrument will switch immediately to battery
power if power from the AC Adapter is interrupted. In this way, measurement can be continued
without losing any data.
If the battery’s remaining charge declines, the instrument will not be able to save or load data.
See “2.2 Installing a Battery Pack” in the Quick Start Manual.
• Set auto saving to [Waveform data].
Waveform data will be saved to an SD Memory Card or a USB Drive while measurement
continues.
You can continue measurement in the event of a power outage by using the Z1007 Battery Pack.
See “3.3 Saving Data” (p. 141).
It is recommended to use the [Binary] (MEM) setting for auto saving.
Files saved using the [Text] (CSV) setting cannot be loaded by the instrument or the Logger
Utility.
Binary data (MEM les) saved using the [Binary] setting can be converted to the text format
using the Logger Utility.
www.GlobalTestSupply.com