Saving Data
Saving settings to the instrument’s internal backup memory
This section describes how to save settings conditions to the instrument’s internal backup memory.
This functionality is convenient when no storage media (SD Memory Card or USB Drive) is
available.
Up to ve groups of settings (labeled No. 1 to No. 5) can be stored in the backup memory.
The settings can be loaded automatically when the instrument is powered on. However, settings
cannot be loaded in the following circumstances:
When the start state retention function is enabled
When the startup auto-function is enabled
When the setting you attempt to load contains the wireless module conguration di󰀨erent from
the one at the time of the shutdown.
> >
2
2
1
1
2
2
3
3
4
4
1
1
1
Under [Auto setup], select whether to load settings conditions from the internal backup
memory when the instrument is powered on.
OFF Do not load settings automatically.
No. 1 to No. 5 Load the settings conditions with the specied group number.
Select A dialog box asking the user to select a group (No. 1 to No. 5) will be displayed.
This function takes priority over settings loaded from the SD Memory Card or USB Drive by the auto setup
function (p. 159).
If [Auto setup] is [OFF], or if no data has been saved in the internal backup memory, the le [STARTUP.SET]
will be loaded from the SD Memory Card or USB Drive.
Saving and Loading Data
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
Storage MediaSD card (up to 32 GB)
Measurement TypesVoltage, Temperature, Resistance, Strain, Pulse, Logic
Sampling Rate1 ms to 1 hour
InterfaceLAN, USB, SD card slot
Power Supply100 to 240 V AC, 12 V DC
Resolution16-bit
Voltage Range±10 mV