333

U8554 Strain Unit
1. General specications
Compatible logger LR8450/LR8450-01 Memory HiLogger
Operating
environment
Indoor use, pollution degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature
and humidity range

Storage temperature
and humidity range

Standards Safety: EN 61010
EMC: EN 61326, Class A
Vibration resistance 
Class 1: Passenger vehicles; conditions: Class A equivalent
(4 h along X-axis and 2 h along Y- and Z-axes at a vibration acceleration of 45 m/s
2

Dimensions 
Weight Approx. 236 g (8.3 oz.)
Product warranty
duration
3 years
Accessories Instruction Manual
Installation screws × 2

2. Input, output, and measurement specications
-1. Basic specications
Number of input
channels
5 (Set voltage or strain for each channel.)
Input terminals 
cover
Set DIP switches according to measurement target.
Measurement target Voltage
Strain Strain gage-type converter
Strain gage
1-gage method (2-wire setup), 1-gage method (3-wire setup),
2-gage method (adjacent sides), 4-gage method
Adaptive gage
resistance
1-gage method, 2-gage method: 120
(external bridge box required for 350
)
4-gage method: 120
to 1 k
Gage factor 
Bridge voltage 2 V ±0.05 V DC
Balance adjustment Method Electronic auto-balancing
Range Voltage: ±20 mV or less (1 mV f.s. to 20 mV f.s. ranges),
±200 mV or less (50 mV f.s. to 200 mV f.s. ranges)
 
ε

ε

ε
f.s. ranges),

ε

ε

ε
f.s. ranges)
Input type 󰀨
A/D resolution 16 bits
Maximum input
voltage
±0.5
V DC

HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
Storage MediaSD card (up to 32 GB)
Measurement TypesVoltage, Temperature, Resistance, Strain, Pulse, Logic
Sampling Rate1 ms to 1 hour
InterfaceLAN, USB, SD card slot
Power Supply100 to 240 V AC, 12 V DC
Resolution16-bit
Voltage Range±10 mV