305
6. Triggers
Trigger method Digital comparison method
Trigger timing Start, stop, or start & stop
Trigger conditions AND/OR operation performed on trigger source, interval trigger, or external trigger
When triggers are disabled, free run
Trigger sources Analog, pulse, logic, waveform calculations, CAN (up to 100)
Trigger types Analog, pulse,
waveform calculations,
CAN
Level triggers
Trigger activated by rising or falling edge at set level
Window triggers
Set by trigger level upper limit and lower limit
Trigger activated when value leaves area or when value
enters area
Logic, CAN Trigger activated when patterns of 1/0/X match (where “X”
indicates either)
Interval triggers Trigger activated for set recording interval after setting days/hours/minutes/seconds
External triggers Trigger activated by rising or falling edge at set level in external input signal
Rising/falling (user-selectable)
Trigger response time With plug-in module(s):
(Recording interval or data refresh interval, whichever is longer) × 2
+ 1 ms + (analog response time)*
1
When using the wireless modules (LR8450-01 only):
(Recording interval or data refresh interval, whichever is longer) × 2
+ (wireless response time)*
2
+ (analog response time)*
1
*2: When the radio-wave state is in good condition, 1s.
Trigger level
resolution
Analog
Pulse Integration 1c, rotational speed 1/n (where n = pulse count per
rotation setting)
Pre-triggers Set day/hours/minutes/seconds.
Can be set during real-time saving.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com