348
Measurement
target
Type Range
Maximum
resolution
Measurable range
Measurement
accuracy
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
100°C f.s. 0.01°C Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than or equal to
100°C
±0.5°C
500°C f.s. 0.05°C Greater than or equal to
±1.4°C
Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than or equal to
500°C
±0.5°C
2000°C f.s. 0.1°C Greater than or equal to
±1.4°C
Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than 500°C
±0.5°C
Greater than or equal to
500°C and less than or
equal to 1350°C
±0.7°C
J 100°C f.s. 0.01°C Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than or equal to
100°C
±0.5°C
500°C f.s. 0.05°C Greater than or equal to
Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than or equal to
500°C
±0.5°C
2000°C f.s. 0.1°C Greater than or equal to
Greater than or equal to
±0.7°C
Greater than or equal to 0°C
and less than or equal to
1200°C
±0.5°C
www.GlobalTestSupply.com