365
-2. Accuracy specications
Accuracy guarantee
conditions
Product warranty duration 1 year
Accuracy guarantee duration
after adjustment made by Hioki
1 year
Accuracy guarantee temperature
and humidity range
Warm-up time At least 30 min. after turning on the module
to 4 kHz.
Measurement range, maximum resolution, measurable range, and measurement accuracy gures
Measurement
target
Range
Maximum
resolution
Measurable range
Measurement
accuracy*
Voltage 1 mV f.s. 50 nV
2 mV f.s. 100 nV ±10 µV
5 mV f.s. 250 nV ±25 µV
10 mV f.s. 500 nV ±50 µV
20 mV f.s. 1 µV ±100 µV
50 mV f.s. 2.5 µV ±250 µV
100 mV f.s. 5 µV ±500 µV
200 mV f.s. 10 µV ±1 mV
Strain 1,000 µ
ε
f.s. 0.05 µ
ε
ε
to 1,000 µ
ε
ε
2,000 µ
ε
f.s. 0.1 µ
ε
ε
to 2,000 µ
ε
±10 µ
ε
5,000 µ
ε
f.s. 0.25 µ
ε
ε
to 5,000 µ
ε
±25 µ
ε
10,000 µ
ε
f.s. 0.5 µ
ε
ε
to 10,000 µ
ε
±50 µ
ε
20,000 µ
ε
f.s. 1 µ
ε
ε
to 20,000 µ
ε
±100 µ
ε
50,000 µ
ε
f.s. 2.5 µ
ε
ε
to 50,000 µ
ε
±250 µ
ε
100,000 µ
ε
f.s. 5 µ
ε
ε
to 100,000 µ
ε
±500 µ
ε
200,000 µ
ε
f.s. 10 µ
ε
ε
to 200,000 µ
ε
±1000 µ
ε
Temperature
characteristics*
Zero position
Voltage: ±1.5 µV/°C
Strain: ±1.5 µ
ε
/°C
Built-in bridge
resistance accuracy
Temperature characteristics: ±2 ppm/°C
Common-mode
rejection ratio
100 dB or greater with a signal source resistance of 300
or less (for 50 Hz/60 Hz input)
Eect of radiated
radio-frequency
electromagnetic eld
ε
f.s. strain range with
Eect of conducted
radio-frequency
electromagnetic eld
ε
*: Does not include built-in bridge resistance tolerance and temperature characteristics.