EasyManuals Logo

Hioki MEMORY HiLOGGER LR8450 User Manual

Hioki MEMORY HiLOGGER LR8450
425 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
42
Conguring Input Channels
When the range is [r/min]
If the time t [s] is less than t
0
(time specied with the smoothing), the displayed rotational speed will
be the actual rotational speed. (However, t
0
is equal to or more than 2 s.)
If an unintentional trigger activates, set the smoothing time at 1 s.
Example when t
0
= 5 s
The rotational speed recorded value will spend t
0
[s] from the start of measurement to increase.
Even if the inputted rotational speed remains constant, the smoothing process will cause the
recorded data to show an increase after the start of measurement until t
0
[s].
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
Storage MediaSD card (up to 32 GB)
Measurement TypesVoltage, Temperature, Resistance, Strain, Pulse, Logic
Sampling Rate1 ms to 1 hour
InterfaceLAN, USB, SD card slot
Power Supply100 to 240 V AC, 12 V DC
Resolution16-bit
Voltage Range±10 mV