42
Conguring Input Channels
When the range is [r/min]
If the time t [s] is less than t
0
(time specied with the smoothing), the displayed rotational speed will
be the actual rotational speed. (However, t
0
is equal to or more than 2 s.)
If an unintentional trigger activates, set the smoothing time at 1 s.
Example when t
0
= 5 s
The rotational speed recorded value will spend t
0
[s] from the start of measurement to increase.
Even if the inputted rotational speed remains constant, the smoothing process will cause the
recorded data to show an increase after the start of measurement until t
0
[s].