114
List of Error Messages
9.4 List of Error Messages
The table below lists messages displayed on the screen in the event of the product or measurement
state error. If repair is necessary, please contact your authorized Hioki distributor or reseller.
RM2611 Electrode Resistance Meter
Display Meaning Solution
NO UNIT The MUX unit has not been
inserted.
Properly insert the MUX unit. If the
error persists, the instrument is
malfunctioning. Please have it repaired.
ERR:090 ROM check sum error. Program ROM checksum error The instrument is malfunctioning. Please
have it repaired.
ERR:091 RAM error. CPU RAM error The instrument is malfunctioning. Please
have it repaired.
ERR:092 Memory access failed.
Main power off, restart
after 10s.
Memory access error Turn off the main power and wait at least
10 s before turning it back on.
ERR:093 Memory read/write
error.
Memory read/write test error The instrument is malfunctioning. Please
have it repaired.
ERR:095 Adjustment data error. Adjustment data error The instrument is malfunctioning. Please
have it repaired.
ERR:096 Backup data error. Conguration backup error The settings have been initialized.
Recongure measurement conditions
and other settings.
ERR:097 Power line detection
error. Select power
line cycle.
Power supply frequency
detection error
The instrument was unable to detect
the power supply frequency. The power
line may be carrying signicant noise.
Either move the instrument away from
the source of noise or afx a ferrite core
or other noise-reducing device to the
power cord.
ERR:098 Blown FUSE or
measurement lead is
broken.
The fuse is blown. Replace the fuse. The fuse is not
customer-replaceable, so please contact
your authorized Hioki distributor or
reseller.
ERR:100 MUX unit error. MUX unit error The instrument is malfunctioning. Please
have it repaired.