Rejecting Error Data during Measurement (Error Data Rejection Function)
4.3 Rejecting Error Data during Measurement (Error
Data Rejection Function)
You can reject error data during measurement. Measurement will continue until a set threshold is
exceeded, even if current errors, over-range errors, or contact errors are encountered.
Rejected error data
Data with current errors
Current channel data with a consistency result of FAIL
Potential data from probes with contact errors
1
2
3
4
5
1
Choose [Settings].
2
Choose [Measurement].
3
Choose one of the following settings for [Error data rejection]:
ON Reject error data.
OFF
Do not reject error data. (Skip step 4 below.) Potential measurement
will stop if a single probe generates a contact error or if a single
potential measurement step generates an error.