Specications
Product components: See p. 5.
For more information about RM9003 Press Unit, RM9004 Test Fixture, and RM9005 Connection Cord
specications, please see those products’ respective instruction manuals.
7.1 RM2611 Electrode Resistance Meter
General specications
Operating environment Indoors, Pollution Degree 2, altitude of up to 2000 m (6562 ft.)
Operating temperature and humidity 0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature and humidity 0°C to 50°C (32°F to 122°F), 80% RH or less (no condensation)
Standards Safety EN61010
EMC EN61326 Class A
Power supply Commercial power
Rated supply voltage: 100 V to 240 V AC (designed to accommodate
voltage uctuations of ±10% relative to the rated supply voltage)
Rated supply frequency: 50 Hz, 60 Hz
Anticipated transient overvoltage: 2500 V
Maximum rated power: 40 VA (not including PC power)
Interfaces • USB port (for connecting the instrument to a PC)
• Test xture connector
• TEMP.SENSOR terminal (for connecting the instrument to the Z2001
Temperature Sensor)
Dimensions Approx. 215W × 80H × 306.5D mm (8.46″W × 3.15″H × 12.07″D)
Mass Approx. 3.4 kg (119.9 oz.)
Product warranty period 3 years
Fuse F1.6AH 250 V ×1
Accessories See p. 5
Options See p. 5
Input, output, and measurement specications
Basic specications
Measured parameters DC resistance and temperature
Automatic measurement is performed via the RM2612 Resistance
Calculation Software. Measurement cannot be performed using the
RM2611 Electrode Resistance Meter alone.
Resistance measurement signal Constant current
Resistance measurement method DC 4-terminal method
Resistance measurement terminal Test xture connector
Resistance measurement range Ranges: 1000 m
Ω
, 10
Ω
, 100
Ω
, 1000
Ω
, 10 k
Ω
, 100 k
Ω
, 1000 k
Ω
For more information about range parameters, see “Table 1. RM2611
Electrode Resistance Meter Measurement Ranges” (p. 76).