EasyManua.ls Logo

Hioki RM2610 - Page 40

Hioki RM2610
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
Setting the Operating Mode
About operating modes
Potential measurement Analysis
Meas. Analysis
Meas. + Analysis
Start of measurement End of measurement
You can choose the operating mode in order to streamline measurement work. When measuring a
small number of electrode sheets, it is recommended to set the mode to [Meas.+Analysis] so that
potential measurement and analysis are performed for each sheet.
When measuring multiple electrode sheets, efciency suffers as the analysis step in
[Meas.+Analysis] results in wait times. By choosing either [Meas.] or [Analysis] as the mode,
you can use the time it takes for analysis results to be generated more effectively by grouping
together analysis of multiple boards, boosting work efciency.
This section describes the workow when measuring multiple electrode sheets.
IMPORTANT
Each mode has its own [File output] setting. Before clicking [Start], check the [File output]
settings.
1
2, 7
6
3, 9
8
Potential measurement
1
Choose [Meas.] under [Mode].
2
Click [OK] to accept the settings.
3
Place an electrode sheet in the RM9003 Press Unit and click [Start] to
perform potential measurement (p. 45).
4
A potential le for analysis use with a lename created by appending a time
stamp to the lename as set in [Path (Analysis results)] will be output. (One
le will be output for each measurement.)
5
Repeat the measurement step for each sheet (Step 3).
Analysis
6
Choose [Analysis] under [Mode].
7
Click [OK] to accept the settings.
8
Select [Analysis result] under [File output].
9
Click [Start] to display a le selection dialog box. Select multiple les output
during [Meas.] operation and click [Open]. (Analysis will be performed one
le at a time, and analysis results will be appended to the analysis results
le.)

Table of Contents

Related product manuals