Brevera Breast Biopsy System User Guide
Chapter 7: Biopsy
Page 86 MAN-04303-002 Revision 006
7.11 How to Change the Biopsy Needle During Troubleshooting
Follow these instructions to change the biopsy needle to troubleshoot issues during setup
and testing.
1. Slide the device driver onto the device driver holder on the console until it clicks.
2. Press the buttons on the sides of the biopsy needle to disengage it from the device
driver. Detach the needle from the driver and place it in a sharps container.
3. Refer to the instructions in Conclude the Procedure on page 84 to disconnect and
remove the biopsy supplies.
4. Press the Check button on the Technologist display to home the device driver.
Warning:
Do not place fingers on the metal parts of the device driver. These parts can
move before a biopsy needle is installed.