Brevera Breast Biopsy System User Guide
Chapter 10: Maintenance, Cleaning, and Disinfecting
Page 100 MAN-04303-002 Revision 006
10.1.3 To Prevent Possible Injury or Equipment Damage
Do not use a corrosive solvent, abrasive detergent, or polish. Select a
cleaning/disinfecting agent that does not damage the plastics, aluminum, or carbon fiber.
Do not use strong detergents, abrasive cleaners, high alcohol concentration, or methanol
at any concentration.
Do not expose equipment parts to steam or high temperature sterilization.
Do not let liquids enter the internal parts of the equipment. Do not directly apply
cleaning sprays or liquids to the equipment. Always use a clean, lint-free cloth and apply
the spray or liquid to the cloth. If liquid enters the system, disconnect the electrical
supply and examine the system before returning it to use.
Caution:
Wrong cleaning methods can damage the equipment, decrease imaging performance,
or increase the risk of electrical shock.
Always follow instructions from the manufacturer of the product you use for cleaning and
disinfection. The instructions include the directions and precautions for the application and
contact time, storage, wash requirements, protective clothing, shelf life, and disposal.
Follow the instructions and use the product in the most safe and effective method.
10.2 Component Specific Cleaning and Disinfection
WARNING!
The console must be disconnected from electrical power source prior to
cleaning. Failure to do so may cause electrical shock or death.
WARNING!
Do not immerse the console in water. Immersion in water will cause
damage and may cause electrical shock or death.
10.2.1 How to Clean and Disinfect the Console Exterior Surfaces
Disconnect the console from the electrical power source.
After each use, clean the console using a soft, damp cloth and mild detergent. Wipe dry.
After cleaning, use a pre-moistened disinfectant wipe or apply a disinfectant solution
to a clean cloth. Wipe all surfaces with the wipe or wet cloth, observing appropriate
contact (wet) times. Refer to General Information on page 99 for recommended
disinfecting solutions.
Do not spray cleaning or disinfectant solutions directly on the console or other parts
of the Brevera breast biopsy system.