Brevera Breast Biopsy System User Guide
Chapter 10: Maintenance, Cleaning, and Disinfecting
MAN-04303-002 Revision 006 Page 101
10.2.2 How to Clean and Disinfect the Device Driver and Remote Control
Disconnect the console from the electrical power source.
After each use, clean the reusable device driver using a soft, damp cloth and mild
detergent. Wipe dry.
After cleaning, use a pre-moistened disinfectant wipe or apply a disinfectant solution
to a clean cloth. Wipe all surfaces with the wipe or wet cloth, observing appropriate
contact (wet) times. Refer to General Information on page 99 for recommended
disinfecting solutions.
Do not spray cleaning or disinfectant solutions directly on the reusable device driver,
the remote control, or other parts of the Brevera breast biopsy system.
Do not immerse the reusable device driver in liquid.
10.2.3 How to Clean and Disinfect the Tissue Filter Drawer
Disconnect the console from the electrical power source.
After each use, clean the tissue filter drawer using a soft, damp cloth and mild
detergent. Wipe dry.
After cleaning, use a pre-moistened disinfectant wipe or apply a disinfectant solution
to a clean cloth. Wipe all surfaces with the wipe or wet cloth, observing appropriate
contact (wet) times. Refer to General Information on page 99 for recommended
disinfecting solutions.
Do not spray cleaning or disinfectant solutions directly in the drawer or other parts
of the Brevera breast biopsy system.
10.2.4 How to Clean the Imaging Display
Disconnect the console from the electrical power source.
Use care when you clean the outer surface of the display area.
Always use a clean, soft, lint-free cloth to clean the display area. Microfiber cloths,
available at most stores, are recommended.
Use a pre-moistened disinfectant wipe or apply disinfectant solution to a clean cloth.
Wipe all surfaces with the pre-moistened wipe or wet cloth, observing appropriate
contact (wet) times. Refer to General Information on page 99 for recommended
disinfecting solutions.
Strong chemicals and abrasives can damage the display window and should never be
used.
Never apply pressure to the display window.
Never use a spray or flow a liquid onto the display.
Never use steel wool.
Never use a sponge with abrasives.