EasyManua.ls Logo

Hologic Brevera - 10.3.6 Preventive Maintenance Schedule for the Service Engineer; Table 5: Service Engineer Preventive Maintenance

Hologic Brevera
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brevera Breast Biopsy System User Guide
Chapter 10: Maintenance, Cleaning, and Disinfecting
Page 108 MAN-04303-002 Revision 006
10.3.6 Preventive Maintenance Schedule for the Service Engineer
Table 5: Service Engineer Preventive Maintenance
Maintenance Task Description
Recommended Frequency
Semiannually
Annually
Obtain previous copy of PM
x
Check with customer for any questions or concerns that
they may have
x
Check error logs
x
Upgrade to current software / firmware configuration if
under warranty / contract
x
Check covers and drawers for mechanical integrity,
missing hardware, clean vents
x
Input wiring inspected for safety and integrity
x
Inspect all ground wiring for integrity
x
Inspect specimen drawer for integrity and sensing
x
Verify operation of all interlocks and bypass functions
x
Verify exposure parameters
x
Check motor accuracy
x
Evaluate image quality
x
Verify completion of Gain Calibration
x
Perform backup timer procedure
x
Clear error log
x
Perform monitor check
x
Test the current leakage on the system
x
Test the resistance on the power cord
x