Brevera Breast Biopsy System User Guide
Chapter 3: Components, Controls, and Indicators
Page 32 MAN-04303-002 Revision 006
WARNING!
The Brevera breast biopsy system console is designed to use the power
cord provided. Do not use a different power cord. Using a different
power cord can create an electrical and fire hazard. DO NOT remove
the ground wire or ground prong from any power plug. DO NOT use
an extension cord with this equipment. An adapter may be necessary
depending on the outlet used.
WARNING!
Make sure that the power cord for the Brevera breast biopsy system
console is in good condition. A damaged power cord can be an
electrical shock hazard. When disconnecting the console from power,
always grasp the plug at the insertion point and pull lightly. NEVER
pull on the cord to unplug the unit.
Warning:
The use of any cables or accessories other than those specified in these
instructions can result in increased emissions or decreased immunity of the
Brevera breast biopsy system.
Caution:
Grounding reliability can only be achieved when this equipment is connected to a
receptacle marked "hospital grade". Check grounding continuity regularly.
3.2.1 Power and Network Connections
1. Put the Brevera breast biopsy system in a location where you can easily access power
connections and network connections.
2. Connect the system AC power cord to the electrical outlet.
Note
Position the console so that you can easily access the AC power cord connection to the
electrical outlet.
3. (Optional) Connect the network cable to the Ethernet connector.