SCPI
COMMAND SUMMARY
l
POWer:OFFSet
<num>~DBllMAXimumlMINimumlUPlDOWN
l
POWer
:
OFFSet?
[MAXimum
I
MINimumI
Sets and queries the power offset,. This function changes mapping of
absolute power parameters on input to and output from the swept
CW generator. Changing this does not, affect the output power of
the swept CW generator. Only the displayed parameters and query
responses are changed. The equation implied by this is:
The entered or displayed power
=
Hardware Power + Offset After
*RST,
the value is 0.
l
POWer:OFFSet:STATe
ONlOFFlllO
a
POWer:OFFSet:STATe?
Queries and turns the power offset off and on. After
*RST,
the
setting is
OFF.
l
POWer:RANGe
<num>[lvl
suffix]
lMAXimumlMINimumlUPlDOWN
l
POWer:RANGe?
Sets and queries the setting of the power meter range. This is used
when the command
POWer:ALC:SOURceis set to
PMETer.
l
POWer:SEARch
ONlOFFlllOlONCE
l
POWer:SEARch?
Sets and queries the power search switch. This has an interaction
with
POWer
: ALC :
STATe
as described below.
POWer:ALC:STATe POWer:SEARch
Power Switch Action
ON
ON
ALC is momentarily closed to
level at the requested power,
and then the modulator is set to
the same voltage in open loop
mode. This repeats automatically
any time that the power level or
frequency is changed.
ON
OFF
Normal mode.
I
ON
ONCE Immediately performs a power
search. This leaves POWer:SEARch
in the ON position.
OFF not applicable
Modulator setting is
explicitly set by user.
l
POWer:SLOPe
<num>CDB/freq
suffix]lMAXimumlMINimumlUPlDOWN
l
POWer:SLOPe?
CMAXimumlMINimuml
Sets and queries the RF slope setting (dB per Hz).
FREQ:MODE Affect on Slope
CW or LIST
Rotates around 0 Hz.
SWEep
or Rotates around the start frequency.
STEP
S-38 Operating and Programming Reference